We analyzed the Smart Groin Trainer device's validity and reliability to measure the isometric hip adduction strength during the adductor squeeze strength test. Fifteen professional soccer players (25.33±4.06 years) and fifteen university students (21.60±1.76 years) participated in this study. All participants performed the squeeze strength test using two portable dynamometers: Smart Groin Trainer and Globus Ergometer. Three maximal isometric hip adduction contractions lasting 5s, interspersed by 3min rest intervals, were performed. Reliability was analyzed with intraclass correlation coefficients, standard error of measurements, and minimal detectable change. The absolute percent error and inter-device accuracy were also analyzed. Correlation analysis assessed the inter-device concurrent validity. The results found no significant differences (p>0.05) between devices in the squeeze strength test values in soccer players and university students. Inter-device comparisons revealed excellent levels of reliability and accuracy in soccer players. Concurrent validity measures revealed strong inter-device relationships in soccer players (r=0.89) and very strong relationships in university students (r=0.99). Intra-device analysis using the Smart Groin Trainer showed excellent relative and absolute reliability in tested soccer players. Our data demonstrated excellent levels of agreement between both dynamometers during the squeeze strength test, suggesting the Smart Groin Trainer as a valid, reliable, and accurate device to measure isometric hip adduction strength.
Key words
- portable dynamometer
- squeeze test
- groin
- accuracy
- measurement error
In sports contexts, hip adduction strength measures are helpful to identify athletes with a high risk of a groin injury and provide an objective result when assessing an athlete's ability to return to play (Thorborg et al., 2011; Tyler et al., 2001). Scientific evidence suggests that groin injuries are common in different sports that require quick changes of movement directions and consequently high levels of accelerations and decelerations, such as in ice hockey and soccer (Esteve et al., 2015; Fulcher et al., 2010; Tyler et al., 2010). Athletes with low levels of isometric hip adduction strength are more prone to develop injuries (Moreno-Pérez et al., 2017; Rodriguez, 2020; Ryan et al., 2014; Tyler et al., 2010). Therefore, measuring hip adduction strength during the season is essential to examine the athlete's needs and prescribe preventive strategies to avoid groin injuries (Tyler et al., 2010).
The force generated by the hip muscles can be measured using different methods, including isokinetic dynamometers, handheld dynamometers, and manual muscle testing (Keep et al., 2016; Meyer et al., 2013). Isokinetic and handheld dynamometers provide quantitative measures, while manual muscle testing offers a qualitative result, which might be more prone to measurement error than the former methods (Keep et al., 2016; Meyer et al., 2013; Nadler et al., 2000). Although isokinetic dynamometers are considered the reference criterion (Stark et al., 2011), they are expensive, non-portable, time-consuming and require large spaces (Martins et al., 2017; Romero-Franco et al., 2019). As an alternative, handheld dynamometers are portable, user-friendly, and cheaper than isokinetic dynamometers (Keep et al., 2016; Martins et al., 2017; Meyer et al., 2013). On the other hand, handheld dynamometers may lead to a lack of stabilization and small substantial changes in the starting position and can generate problems with the placement of the dynamometer (Keep et al., 2016). Nevertheless, several authors have considered handheld dynamometers valid and reliable devices to measure hip adduction strength (Thorborg et al., 2009). Therefore, using a portable dynamometer to measure maximal isometric hip adduction strength seems a valid and reliable option within several sports contexts.
A recent study analyzed the reliability of a portable dynamometer known as a Smart Groin Trainer to measure isometric hip adduction strength in twenty-five university students (Gonçalves et al., 2021). The results demonstrated good inter- and intra-device reliability (intra-class correlation coefficient [ICC] values between 0.89-0.94 and coefficient of variation [CV] values between 3-4%) for the maximal isometric hip adduction strength measured in a supine and a seated position. According to those authors, the Smart Groin Trainer presented reliability levels like those reported in previous studies, making it a reliable device to measure hip adduction strength in sports and rehabilitation contexts. In fact, a previous prospective cohort study with seventy-one elite soccer players used the Smart Groin Trainer device to assess the hip adductor strength and track the incidence of groin injury during a competitive season (Moreno-Pérez et al., 2019). In addition, Moreno-Pérez et al. (2019) reported high reliability and relatively low measurement error when using the Smart Groin Trainer device to assess the hip adduction strength through a soccer-specific test called a squeeze strength test.
Nevertheless, despite the findings mentioned above, to our best knowledge, no study has compared the validity and reliability of the Smart Groin Trainer with another portable dynamometer to measure isometric hip adduction strength. Therefore, comparing the results obtained with the Smart Groin Trainer with another portable dynamometer, such as the Globus Ergometer (Morouço et al., 2014; Neiva et al., 2011), will make it possible to infer the validity and accuracy of this device in different sports contexts.
In this sense, in the current research, we aimed to analyze the intra- and inter-device validity and reliability of the portable Smart Groin Trainer device to measure the isometric hip adduction strength in professional soccer players and university students. We hypothesized that the Smart Groin Trainer device would present high intra- and inter-device validity and reliability to measure the hip adduction strength in professional soccer players and university students.
In this crossover study, we analyzed the intra- and inter-device validity and reliability of the Smart Groin Trainer to measure the isometric hip adduction strength during the squeeze strength test in professional soccer players and university students. Before testing, all participants underwent a familiarization session to learn the adductor squeeze strength test's correct technique. Following the familiarization session, participants performed twice the squeeze strength test using two devices. The first was the Smart Groin Trainer (NeuroExcellence, Braga, Portugal), and the second was the Globus Ergometer (Globus™, Codigné, Italy). The data collection was randomized between participants, and all performed the same number of repetitions. The testing sessions were performed in a university laboratory with a room temperature between 22 and 24º C.
Fifteen male professional soccer players from the 2nd main league of the country (age: 25.33±4.06 years; body mass: 78.71±10.35 kg; body height: 1.81±0.06 m) and fifteen male sport sciences students who regularly performed physical exercise (age: 21.60±1.76 years; body mass: 72.23±8.21 kg; body height: 1.77±0.08 m) volunteered to participate in this study. All participants were free of ankle, knee, or hip injury for the past six months, and none of them presented medical problems in the previous three months. They were instructed to avoid strenuous exercise at least two days before the testing session. All participants received detailed information regarding the study procedures and signed a written informed consent form. The Ethics Committee of the University of Beira Interior approved this study which followed the recommendations of the Declaration of Helsinki.
The Smart Groin Trainer consists of a uniaxial load cell interfaced to a personal laptop through a 12-bit analog-to-digital data acquisition board and custom software. The device measures the applied force up to 4905 N at a sampling frequency of 100Hz (Gonçalves et al., 2021). The Globus Ergometer consists of a data acquisition system that can either export or exhibit the data recorded by a load cell. The load cell measures the exerted force up to 5000N at a sampling frequency of 100Hz (Morouço et al., 2014; Neiva et al., 2011). In both devices, the force data were presented in kg due to the fact that it is easier for coaches/teachers to understand.
Testing sessions started with a general warm-up of 10 min, consisting of running at low to moderate intensity, followed by joint mobilization. Then, all participants performed the adductor squeeze strength test following the procedures described by Moreno-Pérez et al. (2019). They lay down on a mat in a supine position with 45⍛ hip flexion and 90⍛ knee flexion, and the dynamometer was placed between the knees and attached to the thighs with Velcro straps (Figure 1). A specific warm-up of three submaximal isometric hip adduction contractions lasting 5 s, interspersed by 30 s rest intervals, was conducted. After recovery, participants performed three maximal isometric hip adduction contractions ("squeeze the cuff as hard as possible") lasting 5s, followed by 3min rest intervals between the following attempts. Three valid measures with both dynamometers were averaged and registered for further analysis.
Figure 1
Illustration of the adductor squeeze strength test.

The normality was verified using the Shapiro-Wilk test, histograms, and Q-Q plots, while the homogeneity of variances was tested using the Levene's test. Paired samples
We did not observe significant differences between devices in the squeeze strength test's mean values in soccer players and university students (Table 1).
Magnitude of inter-device differences in the measurement of isometric hip adduction strength.
Squeeze Strength Test (kg) | |||||
---|---|---|---|---|---|
Participants | Device | Mean ± SD | Range | ||
Soccer players | Globus Ergometer | 46.18 ± 8.41 | 26.77 to 59.53 | ||
( |
Smart Groin Trainer | 45.69 ± 7.69 | 30.80 to 61.47 | 0.626 | 0.06 (trivial) |
Students | Globus Ergometer | 42.56 ± 12.07 | 22.90 to 62.53 | ||
( |
Smart Groin Trainer | 42.04 ± 12.00 | 24.77 to 61.03 | 0.354 | 0.04 (trivial) |
The effect size was calculated using the Hedge's g formula; p: alpha value between devices.
The Bland-Altman plots are presented in Figure 2.
Figure 2
Bland-Altman plots with 95% limits of agreement (dotted lines) in soccer players (A) and university students (B); Solid lines represent the mean difference between devices (Globus Ergometer vs. Smart Groin Trainer).

Table 2 presents the inter-device reliability and accuracy in the measurement of the isometric hip adduction strength. The ICC data revealed excellent inter-device relative reliability, and CV values moderate and excellent inter-device absolute reliability in soccer players and university students, respectively. The highest MDC was recorded for soccer players. In both cases, the APE was lower than 10%, representing an acceptable measurement error, and the accuracy was higher than 92%, thus revealing a high accuracy level.
Inter-device reliability and accuracy in the measurement of isometric hip adduction strength.
Soccer Players ( |
Students ( |
|
---|---|---|
ICC (95% CI) | 0.94 (0.83-0.98) | 0.99 (0.98-1.00) |
SEM (kg) | 2.71 | 1.46 |
CV (%) | 5.90 | 3.45 |
MDC (kg) | 7.51 | 4.05 |
APE (%) | 7.51 | 4.08 |
Accuracy (%) | 92.49 | 95.92 |
APE: absolute percent error; CI: confidence interval; CV: coefficient of variation; ICC: intra-class correlation coefficient; MDC: minimal detectable change; SEM: standard error of measurement.
The regression models for the two groups are presented in Figure 3. The correlation between devices was strong in soccer players and very strong in university students.
Figure 3
Linear regression models in soccer players (A) and university students (B); CI: confidence interval; r2: coefficient of determination; r: Pearson's correlation coefficient; SEE: standard error of the estimate.

Table 3 presents the intra-device reliability in the measurement of the isometric hip adduction strength. The ICC revealed excellent relative reliability in soccer players, and the CV showed good absolute reliability using both devices. The MDC was higher in the Globus device than the Smart Groin Trainer. The ICC revealed excellent relative reliability in university students using both devices, and the CV presented good absolute reliability. The MDC was higher in the Smart Groin Trainer than in the Globus device.
Intra-device reliability in the measurement of isometric hip adduction strength.
Participants | Device | ICC (95% CI) | SEM (kg) | CV (%) | MDC (kg) |
---|---|---|---|---|---|
Soccer Players | Globus Ergometer | 0.95 (0.88-0.98) | 1.92 | 4.15 | 5.31 |
( |
Smart Groin Trainer | 0.98 (0.94-0.99) | 1.17 | 2.57 | 3.25 |
Students | Globus Ergometer | 0.99 (0.98-1.00) | 0.96 | 2.26 | 2.67 |
( |
Smart Groin Trainer | 0.99 (0.98-1.00) | 1.09 | 2.60 | 3.03 |
CI: confidence interval; CV: coefficient of variation; ICC: intra-class correlation coefficient; MDC: minimal detectable change; SEM: standard error of measurement.
The current research examined the intra-and inter-device validity and reliability of the portable Smart Groin Trainer dynamometer to measure isometric hip adduction strength in professional soccer players and university students during the squeeze test. Our results demonstrated that the Smart Groin Trainer device is valid and reliable for measuring the adductors' strength during the squeeze test, confirming our hypothesis. Therefore, our data suggest the Smart Groin Trainer as a valuable portable device for professional and recreational sports to measure hip adduction strength during the squeeze test.
Previous studies determining the relative reliability of dynamometers to measure hip adduction strength in a supine position observed ICCs from 0.76 to 0.98 when using handheld dynamometers (Kelln et al., (2008), ICC: 0.77 to 0.97; Mentiplay et al., (2015), ICC: 0.84 to 0.98; Thorborg et al., (2009), ICC: 0.76 to 0.98), and ICCs from 0.79 to 0.93 when using an isokinetic dynamometer (Meyer et al., 2013). Therefore, taken together, these results suggest that both devices provide high relative reliability values when measuring hip adduction strength in a supine position. In fact, a previous metanalysis aiming to determine the reliability between isokinetic and handheld dynamometers to measure the force production during different types of joint movements observed ICC values ranging from 0.91 to 0.98 for hip adduction strength measured in a supine position (Chamorro et al., 2017).
Nevertheless, despite the findings mentioned above, to our knowledge, only one study has analyzed the relative reliability of the Smart Groin Trainer to measure hip adduction strength in a supine position (Gonçalves et al., 2021). In that study, the authors observed an ICC value of 0.94 (0.86-0.97) for the maximal isometric hip adduction contraction (N), which is similar to values presented in our research for the intra-device analysis. Our analysis also demonstrated that the Smart Groin Trainer presented high inter-device relative reliability to measure hip adduction strength in professional soccer players and university students. The high ICCs were accompanied by high accuracy values, low APE results, and strong to very strong relationships between devices. Therefore, as stated for handheld and isokinetic dynamometers, a portable and stabilized dynamometer (i.e., that "fixes" or "stabilizes" the load sensor in the assessed muscle region) presents high relative reliability, accuracy, and validity to measure hip adduction strength during the squeeze test.
The intra-device analysis demonstrated small SEM and CV values when using the Smart Groin Trainer. The CV results are similar to those presented in previous research that used the same device to measure the maximal isometric hip adduction contraction (N) in a supine position (Gonçalves et al., 2021). Similar values of absolute reliability were also observed in studies which measured hip adduction strength in a supine position using handheld dynamometers (Kelln et al., 2008; Mentiplay et al., 2015; Thorborg et al., 2009) or an isokinetic dynamometer (Meyer et al., 2013). Besides these results, our study also demonstrates that the SEM and CV data extracted from the comparisons between devices confirm the Smart Groin Trainer's excellent reliability to measure hip adduction strength during the squeeze test. It is essential to notice that the latter test is highly used in soccer contexts due to its high validity and reliability to diagnose groin injuries (Delahunt et al., 2011; Moreno-Pérez et al., 2019). Therefore, using the Smart Groin Trainer during the squeeze strength test might be a valid and reliable strategy to identify athletes with a high risk of groin injuries and monitor changes in adductor strength over the season (Moreno-Pérez et al., 2019).
This study has some limitations that must be considered. Firstly, it is worth mentioning that the study sample does not enable us to generalize the results to other athletic populations. Secondly, the homogeneity of players considering their different playing positions was not taken into account because it could cause increased levels of variability in the final results. Thirdly measuring hip abduction strength would allow us to examine the Smart Groin Trainer's inter-device reliability and analyze the hip adductor to abductor strength ratio. Fourthly, measuring kinetic variables such as the rate of force development over different periods would help us analyze the Smart Groin Trainer's inter-device reliability. Finally, recruiting athletes of various sports with different hip strength levels would be essential to examine the dynamometer's reliability to identify these differences. Therefore, future studies should recruit a broad spectrum of sports athletes to examine the Smart Groin Trainer device's validity and reliability to measure isometric hip adduction and abduction strength and other mechanical variables.
Our results revealed that the Smart Groin Trainer device is valid and accurate to measure isometric hip adduction strength in professional soccer players and university students. This device can be used in clinical and sports contexts to measure isometric hip adduction strength during the squeeze strength test and hip strength changes during the training and rehabilitation process. In addition, the Smart Groin Trainer’s features (portable, small, and easy-to-use) make it a suitable device to be used by coaches and researchers in the sports field.
Figure 1

Figure 2

Figure 3

Inter-device reliability and accuracy in the measurement of isometric hip adduction strength.
Soccer Players ( |
Students ( |
|
---|---|---|
ICC (95% CI) | 0.94 (0.83-0.98) | 0.99 (0.98-1.00) |
SEM (kg) | 2.71 | 1.46 |
CV (%) | 5.90 | 3.45 |
MDC (kg) | 7.51 | 4.05 |
APE (%) | 7.51 | 4.08 |
Accuracy (%) | 92.49 | 95.92 |
Intra-device reliability in the measurement of isometric hip adduction strength.
Participants | Device | ICC (95% CI) | SEM (kg) | CV (%) | MDC (kg) |
---|---|---|---|---|---|
Soccer Players | Globus Ergometer | 0.95 (0.88-0.98) | 1.92 | 4.15 | 5.31 |
( |
Smart Groin Trainer | 0.98 (0.94-0.99) | 1.17 | 2.57 | 3.25 |
Students | Globus Ergometer | 0.99 (0.98-1.00) | 0.96 | 2.26 | 2.67 |
( |
Smart Groin Trainer | 0.99 (0.98-1.00) | 1.09 | 2.60 | 3.03 |
Magnitude of inter-device differences in the measurement of isometric hip adduction strength.
Squeeze Strength Test (kg) | |||||
---|---|---|---|---|---|
Participants | Device | Mean ± SD | Range | ||
Soccer players | Globus Ergometer | 46.18 ± 8.41 | 26.77 to 59.53 | ||
( |
Smart Groin Trainer | 45.69 ± 7.69 | 30.80 to 61.47 | 0.626 | 0.06 (trivial) |
Students | Globus Ergometer | 42.56 ± 12.07 | 22.90 to 62.53 | ||
( |
Smart Groin Trainer | 42.04 ± 12.00 | 24.77 to 61.03 | 0.354 | 0.04 (trivial) |
Inter-Limb Muscle Property Differences in Junior Tennis Players The Pre-Exhaustion Method Does Not Increase Muscle Activity in Target Muscle During Strength Training in Untrained Individuals Neuromuscular Impact of Acute Hypertrophic Resistance Loading With and Without Blood-Flow Restriction Validity and Reliability of the Smart Groin Trainer for Measuring Hip Adduction Strength The Repeated Curve Sprint Test Appears to be an Appropriate Tool for Estimating Anaerobic Fitness in Young Trained Male Futsal Players Training with a Heavy Puck Elicits a Higher Increase of Shooting Speed Than Unloaded Training in Midget Ice Hockey Players Accuracy of Predicting One-Repetition Maximum from Submaximal Velocity in the Barbell Back Squat and Bench Press The Importance of Posture And Body Composition for the Stability and Selected Motor Abilities of Professional Handball Players Influence of Tactical Behaviour on Running Performance in the Three Most Successful Soccer Teams During the Competitive Season of the Spanish First Division Post-Activation Performance Enhancement in Sprinters: Effects of Hard Versus Sand Surfaces Cognitive Factors in Elite Handball: Do Players’ Positions Determine their Cognitive Processes? Effects of Postactivation Performance Enhancement on the Vertical Jump in High-Level Volleyball Athletes On-Court Change of Direction Test: An Effective Approach to Assess COD Performance in Badminton Players Movement Patterns of Polish National Paralympic Team Wheelchair Fencers with Regard to Muscle Activity and Co-Activation Time The Creation of Goal-Scoring Opportunities at the 2019 FIFA Women’s World Cup Assessing the Sprint Force-Velocity Profile in International Football Players with Cerebral Palsy: Validity, Reliability and Sport Class’ Profiles Handcycling Training in Men with Spinal Cord Injury Increases Tolerance to High Intensity Exercise Hand Grip Strength vs. Locomotor Efficiency in Sitting Volleyball Players Evidence‐Based Recovery in Soccer – Low‐Effort Approaches for Practitioners Anaerobic Variables as Specific Determinants of Functional Classification in Wheelchair Basketball Addition of in‐Play Cooling Breaks During Intermittent Exercise while Wearing Lacrosse Uniforms in the Heat Attenuates Increases in Rectal Temperature The Effects of Preferred Music and Its Timing on Performance, Pacing, and Psychophysiological Responses During the 6‐min Test Prevalence of Dehydration and the Relationship with Fluid Intake and Self‐Assessment of Hydration Status in Czech First League Soccer Players A Study Comparing Gait and Lower Limb Muscle Activity During Aquatic Treadmill Running With Different Water Depth and Land Treadmill Running Changes in Heart Rate Variability and Post‐Exercise Blood Pressure from Manipulating Rest Intervals Between Sets of Resistance Training