Calibration of NMR Receiver using Spectrometer Characteristics

Published Online: 26 Oct 2021
Page range: 205 - 208
Received: 15 Aug 2021
Accepted: 14 Oct 2021
Journal Details
License
Format
Journal
First Published
07 Mar 2008
Publication timeframe
6 times per year
Languages
English
Abstract

This article describes the measurement of the relation between input and output signals using two techniques: with a signal generator and with the thermal noise of a known resistance. Each of the techniques has its advantages and disadvantages. Both methods are tested and the results are compared. The input signal of the receiver is known in volts, while the output signal is in ADC (analogue-to-digital converter) units. It is the main difference versus the gain. Knowledge of the relation enables recalculation of the output signal into the input of the receiver or vice versa. It is important in some experiments. The method with the harmonic signal requires a suitable NMR spectroscopic console, generator of the harmonic signal and an attenuator, the method with the noise requires only the NMR console. It indicates that both methods are simple and cheap. The measured data are processed on a standard PC using common programs.

Keywords

[1] Andris, P., Emery, E.F., Frollo, I. (2019). Analysis of NMR spectrometer receiver noise figure. Mathematical Problems in Engineering, 2019, 1083706. https://doi.org/10.1155/2019/1083706.10.1155/2019/1083706 Search in Google Scholar

[2] Schiek, B., Siweris, H.J. (1990). Rauschen in Hochfrequenzschaltungen [Noises in RF Circuits]. Hüthig, ISBN 978-3778520079. (in German) Search in Google Scholar

[3] Žalud, V., Kulešov, V.N. (1980). Polovodičové obvody s malým šumem [Semiconductor Circuits with Low Noise]. Prague, Czech Republic: SNTL, ISBN 04-528-80. (in Czech) Search in Google Scholar

[4] Hoult, D.I., Richards, R.E. (1976). The signal-to-noise ratio of the nuclear magnetic resonance experiment. Journal of Magnetic Resonance, 24 (1), 71-85. https://doi.org/10.1016/0022-2364(76)90233-X.10.1016/0022-2364(76)90233-X Search in Google Scholar

[5] Hoult, D.I., Lauterbur, P.C. (1979). The sensitivity of the zeugmatographic experiment involving human samples. Journal of Magnetic Resonance, 34 (2), 425-433. https://doi.org/10.1016/0022-2364(79)90019-2.10.1016/0022-2364(79)90019-2 Search in Google Scholar

[6] Raad, A., Darrasse, L. (1992). Optimization of NMR bandwidth by inductive coupling. Magnetic Resonance Imaging, 10 (1), 55-65. https://doi.org/10.1016/0730-725x(92)90373-8.10.1016/0730-725X(92)90373-8 Search in Google Scholar

[7] Décorps, M., Blondet, P., Reutenauer, H., Albrand, J.P., Remy, C. (1985). An inductively coupled, series-tuned NMR probe. Journal of Magnetic Resonance, 65 (1), 100-109. https://doi.org/10.1016/0022-2364(85)90378-6.10.1016/0022-2364(85)90378-6 Search in Google Scholar

[8] Vergara, Gomez, T.S., Dubois, M., Glybovski, S., Larrat, B., De Rosny, J., Rockstuhl, C., Bernard, M., Abdeddaim, R., Enoch, S., Kober, F. (2019). Wireless coils based on resonant and nonresonant coupled-wire structure for small animal multinuclear imaging. NMR in Biomedicine, 32 (5), e4079. https://doi.org/10.1002/nbm.4079.10.1002/nbm.4079 Search in Google Scholar

[9] Qian, Ch., Duan, Q., Dodd, S., Koretsky, A., Murphy-Boesch, J. (2016). Sensitivity enhancement of an inductively coupled local detector using a HEMT-based current amplifier. Magnetic Resonance in Medicine, 75 (6), 2573-2578. https://doi.org/10.1002/mrm.25850.10.1002/mrm.25850 Search in Google Scholar

[10] Weis, J., Ericsson, A., Hemmingsson, A. (1999). Chemical shift artifact-free microscopy: Spectroscopic microimaging of the human skin. Magnetic Resonance in Medicine, 41 (5), 904-908. https://doi.org/10.1002/(SICI)1522-2594(199905)41:5%3C904::AIDMRM8%3E3.0.CO;2-4. Search in Google Scholar

[11] Marcon, P., Bartusek, K., Dokoupil, Z., Gescheidtova, E. (2012). Diffusion MRI: Mitigation of magnetic field inhomogeneities. Measurement Science Review, 12 (5), 205-212. https://sciendo.com/de/article/10.2478/v10048-012-0031-8. Search in Google Scholar

[12] Bartusek, K., Dokoupil, Z., Gescheidtova, E. (2007). Mapping of magnetic field around small coil using the magnetic resonance method. Measurement Science and Technology, 18 (7), 2223-2230. https://doi.org/10.1088/0957-0233/18/7/056.10.1088/0957-0233/18/7/056 Search in Google Scholar

[13] Nespor, D., Bartusek, K., Dokoupil, Z. (2014). Comparing saddle, slotted-tube and parallel-plate coils for Magnetic Resonance Imaging. Measurement Science Review, 14 (3), 171-176. https://doi.org/10.2478/msr-2014-0023.10.2478/msr-2014-0023 Search in Google Scholar

[14] Latta, P., Gruwel, M.L., Volotovskyy, V., Weber, M.H., Tomanek, B. (2007). Simple phase method for measurement of magnetic field gradient waveforms. Magnetic Resonance Imaging, 25 (9), 1272-1276. https://doi.org/10.1016/j.mri.2007.02.002.10.1016/j.mri.2007.02.002 Search in Google Scholar

[15] Latta, P., Gruwel, M.L., Volotovskyy, V., Weber, M.H., Tomanek, B. (2008). Single-point imaging with a variable phase encoding interval. Magnetic Resonance Imaging, 26 (1), 109-116. https://doi.org/10.1016/j.mri.2007.05.004.10.1016/j.mri.2007.05.004 Search in Google Scholar

[16] Wimmer, G., Witkovský, V., Duby, T. (2000). Proper rounding of the measurement results under normality assumptions. Measurement Science and Technology, 11 (12), 1659-1665. https://doi.org/10.1088/0957-0233/11/12/302.10.1088/0957-0233/11/12/302 Search in Google Scholar

[17] Witkovsky, V., Frollo, I. (2020). Measurement science is the science of sciences - there is no science without measurement. Measurement Science Review, 20 (1), 1-5. https://doi.org/10.2478/msr-2020-0001.10.2478/msr-2020-0001 Search in Google Scholar

Recommended articles from Trend MD

Plan your remote conference with Sciendo