Efficacy of breast shielding during head computed tomography examination


Cite

Introduction

Computed tomography (CT) is a valuable diagnostic tool for many conditions.1 According to the International Atomic Energy Agency (IAEA) an individual receives 2.4 mSv of the total annual dose of natural radiation2 and the typical effective radiation dose received during a standard CT scan ranges from 1 to 14 mSv.3 Head CT is one of the most frequently performed examinations 4,5 and accounts for 50% of all CT scans and 25% of the collective radiation dose from CT.5 High effective doses received during CT examinations raise concerns1,6, so an effort to minimize it is crucial.1 Due to concerns over radiation exposure, the very valuable principle ‘‘as low as reasonably achievable’’ (ALARA) was established to stress the importance of minimizing radiation exposure.7 Many CT shielding techniques have been developed as a result, among them radiation protection shields, which are one of the simplest mechanisms by which radiation dose can be reduced.3

The use of lead shielding is a long-accepted method to minimize the dose from scatter radiation in radiographic procedures.1,8 In head CT, the head as the main target under investigation is included in the primary beam, while radiosensitive organs such as the breasts are located outside the primary beam. Radiosensitive structures that lie in the primary beam cannot be easily protected, while the organs outside the primary beam, which are mainly exposed to scattered radiation, can be easily protected against it. Scattered radiation distributes caudally through patient’s neck thereby inevitably exposing the breasts. However, scattered radiation produced in gantry and patient’s head reaching the breasts from outside can be effectively diminished by the lead shielding.8 Radiation dose to the breasts is significant enough to be a matter of concern9 regardless of all the current dose reduction techniques.5

Female breast tissue is highly sensitive to radiation’s carcinogenic effects. Its tissue weighting factor has risen from 0.05 to 0.12.7 The exposure of breasts to ionising radiation is a strong risk factor for breast cancer. The study by Preston et al.10 showed that the increased risk is directly proportional to the radiation dose received and inversely related to age at irradiation. The use of radiation shields, if applied properly, is one of the simplest implementations by which radiation dose can be diminished without the negative impact on diagnostic image quality.3

The benefit of shielding to radiosensitive organs in many diagnostic procedures has been acknowledged in many studies.1,5,8,11, 12, 13 Brnić et al.8 recommend lead shielding of the breasts during head CT since a recognizable reduction (57%) of the exposure was found. The study by Williams and Adams5 also discovered a significant dose reduction (58–47%) with a lead bib protecting the thyroid. Beaconsfield et al.11 also claim a significant dose reduction (76%) during head CT due to lead shielding of breasts. However, Brnić et al.8 measured the exposure with thermoluminescent dosimeters (TLD) during shielding of only one breast while the other remained unshielded, the study by Williams and Adams5 performed measurements while investigating the benefit of thyroid shielding on an anthropomorphic phantom and Beaconsfield et al.11 have performed only 20 measurements with the use of breast shielding during axial head CT examination.

For the above-mentioned reasons and given the high frequency of head CT scans and high sensitivity of breasts to ionising radiation, the purpose of this study was to determine the effectiveness of breast shielding during the routine helical head CT scan in real clinical practice with an electronic dosimeter and a lead apron of 0.5 mm lead equivalent, which is the most commonly used protective shield in Slovenia. Lead shielding is not routinely used in the breast area during head CT examination so the results could clarify the dilemma whether the shielding is beneficial in protecting the breasts. Possible changes in clinical practice might be expected in the future.

Patients and methods

The study was performed in two major hospitals in Slovenia. The sample size was calculated based on the study performed by Brnić et al.8, with the G*Power 3.1 (Allgemeine Psychologie und Arbeitspsychologie, University Dusseldorf) programme. 120 female patients referred to a spiral head CT scan participated in our study (age range 23–92 years with average age of 67.9 ± 16.7 years). 60 measurements were conducted in Hospital A on Siemens Somatom Definition (Siemens Medical Systems, Germany) CT unit and 60 measurements in Hospital B with Toshiba Aquilion 64 slice (Toshiba Medical Systems, Otawara, Japan) CT unit. All results of the quality assurance tests for both CT units, performed prior to the study by the medical physicist, were under the remedial level. Half of the measurements in each hospital (n = 30) were performed with the use of lead shielding and half of the measurements (n = 30) without the lead shielding. Randomized sampling was used in this study.

The Unfors EDD 30 (Unfors, Sweden) unit was used to measure the skin dose at the surface of the breast in real time during CT of the head. According to the manufacturer, the dosimeter offers a dose range from 1 nGy to 9999 Gy with the start trigger level at 0.05 mSv/h. The manufacturer also claims that the EDD 30’s inaccuracy is ± 6% at calibration point with 80 kVp, 2 mm of Cu and 4 mm of Al. The dosimeter was tested and showed the best and consistent response of the dosimeter when it was in alignment to the horizontal plane and the cable of the dosimeter was positioned away from the source of scattered radiation (not contained in the scanning area). The dose was read on the display.

Acquisition parameters are shown in Table 1. Same protocols as used in clinical practice in each hospital were used during the study.

Acquisition parameters

HospitalAB
Pitch0.550.656
Rotation time (s)10.6
kV120120
mA480200

The sensor of the dosimeter was attached to the centre of the breast in the alternating order, in a way that the dosimeter was always in alignment with the examination table to ensure minimal effect of the angulation dependence of the dosimeter. The breast area was wrapped with lead apron (Mavig, Germany) of 0.5 mm equivalent lead thickness as tightly as possible around the patient (shoulders were not draped). The example of the shielding position is presented on Figure 1 using an anthropomorphic phantom.

Figure 1

Presentation of the shielding position used in the study.

Our study was approved by the hospital’s Ethics Committee and by the National Medical Ethics Committee prior to the initiation. A written informed consent was obtained from all patients. All the measurements were processed with the IBM SPSS STATISTICS version 25 (IBM, USA). The Shapiro-Wilk test was used to check the normal distribution of the sample. A student T-test was performed to determine the statistical significance of the difference of the skin dose between the shielded and unshielded breasts in each hospital. A significance of p < 0.05 was used for all the tests.

Results

In Hospital A the mean skin dose to the breast was 338.2 ± 43.7 μGy without the lead shielding and 64.3 ± 18.8 μGy when the lead shielding was used. The difference between the doses was statistically significant (p < 0.001) with the dose reduction of approximately 81%. In Hospital B, the mean skin dose to the unprotected breasts was 253.1 ± 35.1 μGy compared to 65.3 ± 16.9 μGy to the breasts protected with the lead shielding. The dose reduction of 74% was statistically different (p < 0.001). Table 2 and Figure 2 summarize the results.

Mean measured skin dose to the breasts in both investigated hospitals

Hospital AHospital B
unshieldedshieldedunshieldedshielded
Mean skin dose to breast (μGy)338.2 ± 43.764.3 ± 18.8253.1 ± 35.165.3 ± 16.9
Dose reduction81%74%

Figure 2

Mean skin dose to breast in both hospitals with and without the shielding.

Discussion

This study provides evidence, that lead shielding significantly reduced the amount of scatter radiation to the breasts in both Hospitals.

Radiosensitive structures that lie in the primary field cannot be easily protected, meanwhile the structures outside the CT scanning plane, such as the breasts during the head CT examination, can be easily protected against the scatter radiation. During the head CT examination the use of lead shielding can significantly reduce the exposure from scattered radiation generated in patient’s head and in gantry.8,14

Our results showed that the breast exposure in Hospital A and B was reduced by 81% and 74%, respectively, with the use of 0.5 mm lead shielding. Nevertheless, a noticeably higher mean skin dose to breasts without the shielding in Hospital A, compared to Hospital B has been detected. This difference might be the result of the higher mA setting, longer scan lengths (were not noted) and the angular dependence of the dosimeter. Further investigation could be made.

The study from Beaconsfield et al.11 reported an average 76% dose reduction using a 0.5 mm lead equivalent thickness, which is similar to our study. They used the axial imaging technique of the head CT examination and not the spiral, as did we in our study. The axial imaging technique was also used in the study by Brnić et al.8, which reported lower reduction in dose; 57%. However, their study used a 0.35 mm lead shield, and only one breast was shielded during the measurements, so a lower dose reduction is expected. However, Williams and Adams5 did use the same CT imaging technique – the spiral scanning. They have investigated the impact of thyroid shielding on dose to the patient during head CT. They have measured a 58% dose reduction on the thyroid during the spiral head CT examination. The thyroid, however, cannot be protected from the X-rays scattered from within the patient.11 All mentioned studies used thermo-luminescent dosimeters to measure the radiation dose. In contrary, our study used an electronic dosimeter so the skin dose on the breasts was read out on the display as soon as the scanning was finished.

The International Commission on Radiological Protection (ICRP) has published new tissue weighting factors for the breast and other organs in 2007. The breast tissue weighting factor was increased from 0.05 to 0.12.7 One of the strongest risk factors for breast cancer is the radiation exposure at a young age.10,15 According to the ALARA principle all effort should be made to reduce irradiation of the breasts using shielding. In addition to breast shielding some other radiosensitive tissues will also be protected against the external scatter radiation, particularly bone marrow (42% of bone marrow is found in the thorax).8,11 Based on the ICRP1037, bone marrow is a radiosensitive organ with the same tissue weighting factor as the breasts – 0.12, however, the effect of shielding on the bone marrow was not investigated in our study.

The study was conducted on female patients and not on an anthropomorphic phantom as the study by Williams and Adams5, so the measurements were performed on real body tissue in real clinical circumstances. In addition, patient positioning and diverse body geometry of the patients influenced the shielding possibilities. Shielding in the present study was wrapped around the chest of the patient so both breasts were shielded simultaneously and not one breast only like in the study by Brnić et al.8 Wrapped-around shielding is also the routine way of shielding in CT diagnostics’ clinical practice, and not covering only one radiosensitive organ. Since shielded and unshielded measurements on the same patient could not be performed, a large sample was necessary for statistical accuracy. In total, 60 measurements of radiation dose to the breasts under the shielding were performed compared to the 20 measurements performed in the study by Beaconsfield et al.11 In the present study the measurements were performed in two different institutions on two different CT units with different exposure parameters. All the above contributes to the additional value of the present work.

Our study has proven that the use of lead shielding significantly reduced the skin dose to the breasts. Nevertheless, special attention still needs to be payed to the positioning of the apron. The lead shielding must be positioned so it does not enter the primary field. If the shielding enters the primary field of imaging, it would produce significant image artefacts or, if the automatic exposure control is used, the system would drastically increase the tube output in order to compensate the absorption in lead. Caution must also be taken to wrap the shielding around the patient as tightly as possible, so there is no air gap between the chest and the shielding.

This study was performed with a lead shield, which is the most commonly used form of shielding in Slovenia. New shielding materials are emerging, among them the Antimony-Bismuth shields, which are positioned in the primary field and only partially block the beam. Further investigation could be performed. Nevertheless, the breasts are located outside the primary field so the lead shielding can be applied to reduce the scattered radiation to the breasts.

This study showed that the lead shielding significantly reduced the amount of scatter radiation irradiating the breasts. The dose was reduced by a minimum of 74% and a maximum of 81%. Head CT is one of the most frequently performed CT examinations and the breasts are one of the most radiosensitive organs so patients undergoing a head CT scan will benefit significantly when lead shielding is used. Our work could be continued by investigating the efficiency of the newly emerging shielding materials, its efficiency in correlation to material thickness and the frequency of using the shielding in different departments.